Thermo Fisher Scientific C101 Colorimeter User Manual

Page 31

Advertising
background image

Measurement: Chlorine Dioxide

C103 Colorimeter

27

6.

Subtract the result obtained in step 5 from the result obtained in step
1 to give the true result.

7.

Correcting for dilution:

To correct for any reagent added to the

sample to neutralise it, a dilution correction factor must be used:

Final Volume ÷ Initial Sample Volume = Correction Factor

The final volume can be calculated by adding together the initial
sample volume and the volume of neutralising reagent used:

Final Volume = Initial Sample Volume + Volume of Neutralising
Reagent.

The result from a test can then be multiplied by the Correction Factor
to give the true result.

If reagents are added from the supplied dropper bottles, then 25
drops is equal to 1ml.

Example:

50ml sample is neutralised using 50 drops of 1N Sulphuric Acid, and
each drop is 0.04 ml.

Initial Sample Volume = 50ml

Volume of Neutralising Reagent added = 50drops X 0.04 ml= 2ml

Final Volume = 50ml + 2ml = 52ml

Correction Factor = 52/50 = 1.04

A test result of 1.00ppm would then become 1.00ppm X 1.04 =
1.04ppm

2.7.4 Accuracy Check / User Calibration

Producing chlorine standards is a difficult procedure that should only be
attempted by trained laboratory staff. Details of preparation techniques are
not provided in this document.

The prepared chlorine standard may be used to check the accuracy of the
meter or to re-calibrate the meter. In almost all cases it is recommended
to use the default factory calibration.

To check the accuracy of the meter:

Prepare a standard solution of

known concentration within the range of the instrument. It is advisable to
prepare a standard near to the concentration value usually tested with the
meter. Use this standard in place of the sample water in the above test
method. Repeat the measurement 5-7 times with different sachets and

Advertising