Thermo Fisher Scientific Alpha COND 560 User Manual

Page 11

Advertising
background image

5

2.3 Connecting

Peripherals

2.3.1 Connection

Terminals

Remove Back Cover:

Remove the screws from the four corners at the back of the conductivity
controller. Remove the back cover. The connectors are exposed on the
back PCBA as shown in the Figure 1 below.

Connectors:

J11 – 24V DC power connector

J8 – 9V DC power connector (1.2mm diameter)

J9 – Relay connections

J10 - Conductivity electrode & Temperature probe connections (wiring
has to be done in the detachable connector

J10

J8

REL 1

REL 2

J9

9V DC Power

Screw

Figure 1: Outer Side of Back PCBA

J11 Connections
1. +ve
2. Gnd

J10 Connections
1.

CON Low Current

2.

CON Low Voltage

3. CON

High

Voltage

4. CON

High

Current

5.

Pt 100 Compensate

6.

Pt 100 Sense

7.

Pt 100 GND

8. (No

connection)


J9 Connections
Relay 1 & 2

1)+ve
2)Gnd