5.3 Temperature

Settings

Automatic Temperature Compensation (ATC):

The conductivity value of a sample is affected by temperature. Use ATC feature of
the controller to compensate for conductivity changes when the temperature of the
sample or process liquid fluctuates. Connect a separate temperature probe
(Pt100/pt1000) to the controller if an integrated conductivity/Temperature probe is
not available.

Manual Temperature Compensation (MTC):

Set the controller to MTC (disable ATC) when the temperature of sample or process
liquid is constant and a temperature probe is not available. If you disable ATC, the
controller allows you to set your process temperature (P.ºC) and calibration
temperature (C.ºC). This allows calibration at a different temperature other than the
process temperature. Example: Setting calibration temperature of 25°C lets you
calibrate using standard solutions at 25°C, even if your process temperature is
different from 25°C.

ENT

°F

SETUP

HOLD

1

SETUP

HOLD

2

°C

SETUP

HOLD

2

ENT

ENT

SETUP

HOLD

°C

3

SETUP

HOLD

°C

3

ENT

SETUP

HOLD

°C

ATC

SETUP

HOLD

°C

ATC

5

4

SETUP

HOLD

ENT

6

SETUP

HOLD

7

ENT

ENT

ENT

1

1

From measurement mode press ENT key to enter setup mode as
described in section 5.1. The LCD shows the first screen of setup mode
(rAng). Press

▲ or ▼ key to select Temperature settings screen (SET

ºCF).

Press ENT key to access temperature settings (SET ºCF).

2

Selecting unit of measurement for temperature: The upper display
shows ‘Unit’ and the lower display shows the last configured unit of
measurement for temperature.

Press

▲ or ▼ key to select the desired units for temperature (ºC or

ºF). Press ENT key to confirm your selection.

3

Enable/disable ATC: The lower display shows ‘AtC’ and the upper
display shows the last configured ATC selection (‘On’ or ‘OFF’).