6

T E C H N I CAL S P E CI F I C AT I O NS

General Specification

Conductivity

Measuring Range:

Resolution:

0.000 - 2.000 µS/cm

0.001 µS/cm

0.00 - 20.00 µS/cm

0.01 µS/cm

0.0 - 200.0 µS/cm

0.1 µS/cm

0- 2000 µS/cm

1 µS/cm

0.00 - 20.00 mS/cm

0.01 mS/cm

0.0 - 200.0 mS/cm

0.1 mS/cm

Relative accuracy

± 1% of full scale reading

System accuracy

The effective measuring range and the
total system accuracy depend on the
sensor used with the controller. Please
refer respective sensor documentation.

Temperature

Measuring range

-10.0 to +125.0

o

C

+14.0 to +257.0

o

F

Resolution 0.1

o

C / 0.1

o

F

Relative accuracy

±0.5

o

C / ±1.0

o

F

Sensor

2 wire / 3 wire - Pt100 or Pt1000 (Jumper
Selectable)

Compensation

Automatic or Manual (reference at 25

o

C)

Calibration

Conductivity

Number of calibration points

1 point

Calibration window

±40% of default reading

Calibration standard solution
window

10% to 100% of the selected measuring
range

Temperature

Offset Adjustment

± 10 ˚C/ ± 18 ˚F

Controller

Output 2

SPST

Relays

Relay 1/Relay 2 function (for
Set Point 1 (SP1) & Set
Point(SP2))

LO (Relay is active when measured value
< Set point)
HI (Relay is active when measured value >
Set point)

Hysteresis window

0% to 10 % of selected measuring range

On-delay & Off-delay time

0 to 2000 seconds

Display

LCD

7 segments display with symbols for status
information

Back light

On/Off selectable