5 controller settings – Thermo Fisher Scientific Alpha COND 560 User Manual

Page 30

Advertising
background image

24

5.5

Controller Settings

The controller settings allow you to set the relay type and relay status of the
controller.

ENT

SETUP

HOLD

2

ENT

SETUP

HOLD

3

SETUP

HOLD

1

SETUP

HOLD

2

ENT

SETUP

HOLD

3

ENT

ENT

1

1

From measurement mode press ENT key to enter setup mode as
described in section 5.1. The LCD shows the first screen of setup
mode (rAng). Press

▲ or ▼ key to select controller settings screen

(Cntr).

Press ENT key to access controller settings (Cntr).

2

Select the controller type: The lower display shows ‘tyP’. The upper
display shows the last configured controller type. (‘OFF’ or ‘L.Ct’).
Press

▲ or ▼ key to select the desired controller type.

OFF :

Controller Off
Selecting ‘OFF’, the conductivity controller stops
functioning as a controller by preventing relay functions.
Once the controller type is set to ‘OFF’, the setup mode
does not show you the Relay 1/Relay 2 settings.

L.Ct:

Limit value control (Controller On)
Selecting ‘L.Ct’, the conductivity controller is set to function
as a controller. When the controller type is set to ‘L.Ct’, the
Relay 1/Relay 2 settings are available in the setup mode
for you to configure relay operating parameters.

Press ENT key to confirm your selection.