4 relay 1/relay 2 settings – Thermo Fisher Scientific Alpha COND 560 User Manual

Page 28

Advertising
background image

22

5.4

Relay 1/Relay 2 Settings

NOTE:

If controller type is set to ‘OFF’ (Refer section 5.5 for controller
settings), the ‘Relay 1/Relay 2 Setting’ is not available.

The SP1 settings determine the operating parameters for Relay 1; while SP2
settings define the operating parameters for Relay 2. Since these two settings have
identical setup parameters, only SP1 is described here.

Note that when you change measuring range of the controller, the values you have
set for SP1 & SP2 will not be affected. However, the unit of measurement &
resolution displayed in LCD for SP1 & SP2 will change according to the newly
selected measuring range.

SETUP

HOLD

1

ENT

SETUP

HOLD

2

ENT

SETUP

HOLD

SETUP

HOLD

3

3

ENT

ENT

SETUP

HOLD

4

ENT

SETUP

HOLD

SETUP

HOLD

ENT

ENT

5

6

µS

µS

µS

µS

1

1

From measurement mode press ENT key to enter setup mode as
described in section 5.1. The LCD shows the first screen of setup mode
(rAng). Press

▲ or ▼ key to select relay 1 (Set point 1) settings

screen (SP1).

Press ENT key to access relay 1 settings (SP1).

2

Setting set point value: The upper display shows the last configured
set point value, otherwise the default value 100 µS. Press

▲ or ▼ key

to enter the value for set point 1, at which control function of the
conductivity controller should activate. Press ENT key to confirm your
set point value.

3

Select relay function: The upper display shows the last configured
relay function (LO = Low or HI=High). Press

▲ or ▼ key to select the

desired function.

Press ENT key to confirm your selection.