5.7 Configuration

Settings

Configuration settings let you to enable/disable backlight of the LCD, select
between 2-cell & 4-cell and reset the controller to factory defaults.

ENT

SETUP

HOLD

2

SETUP

HOLD

2

ENT

ENT

SETUP

HOLD

3

SETUP

HOLD

3

ENT

SETUP

HOLD

SETUP

HOLD

5

4

ENT

ENT

SETUP

HOLD

ENT

ENT

SETUP

HOLD

5

SETUP

HOLD

5

ENT

1

MEAS

°C

ATC

READY

µS

This screen appears only

when 2-cell is selected

and measuring range is set to

0.0 to 200.0 mS/cm

R.

1

1

From measurement mode press ENT key to enter setup mode as
described in section 5.1. The LCD shows the first screen of setup mode
(rAng). Press

▲ or ▼ key to select configuration settings screen

(Cnfg).

Press ENT key to access configuration settings (Cnfg).

2

Enable/disable LCD backlight: The upper display shows ‘LitE’. The
lower display shows last configured backlight selection (‘On’ or ‘OFF’).

Press

▲ or ▼ key to enable (LitE On) or disable (LitE OFF) backlight