8 TRO UBLE SHOO TIN G

Problem

Cause

Solution

Power on, but no
display

a) Loose

connections

b) Incorrect

output

voltage

of the power adaptor

c)

Cables not in correct
polarity (+ and – position)

a)

Ensure cables make good
contact

b)

Use an power adaptor with
specified output voltage

c)

Re-wire loop cables with
correct polarity

Unstable conductivity
reading

a)

Air bubbles in the probe

b) Dirty

probe

c)

Probe not deep enough
in sample

a)

Tap probe to remove air
bubbles

b)

Clean probe and
recalibrate

c)

Ensure that the sample
entirely covers the sensor
of the probe

Oscillating
temperature readings

a) Electrical

noise

interference

a)

Ensure shield wire is
properly connected to pin 7

Slow response

a) Dirty / Oily probe

a) Clean probe

Blinking ATC

a)

No temperature probe
connection during ATC
mode

a) Ensure

temperature

sensing cable makes good
contact

Blinking electrode
annunciator

a)

Error in calibration

b)

Conductivity value of the
calibration standard is not
within acceptable range

a) Ensure

calibration

standard solution is not
contaminated. Ensure
electrode is clean

b) Ensure

calibration

standard is within 10 to
100% of full-scale of the
selected measuring range

Or (Conductivity)

a)

Conductivity reading is
above the selected
measuring range

a)

Ensure correct conductivity
measurement range is
selected

Or/Ur (Temperature)

a)

Temperature probe is not
connected when ATC
enabled

b) Temperature

value

exceeds above/ below
the specified range

a)

Ensure electrode makes
good contact with
controller

b) Ensure

temperature

calibration is done
properly. Ensure process
temperature is within the
specified range