32

Electrical Data and Connections

Conductivity Input

Screw Terminal (3.5mm pitch)

Connection terminal

8-pin, 2-pin & 5-pin terminal blocks, DC
power connector (1.2mm diameter)

EMC Specifications

Emission

EN 61 36

Susceptibility

EN 61326

Other

Power Input

Universal +9 V DC Adaptor and 24V DC
Supply

Dimensions (W x H x D)

96mm x 96mm x 66 mm

Weight (Estimated)

210g

Ambient Temp. operating range

0 to 40

o

C

Maximum Relative Humidity

80% up to 31

o

C decreasing linearly to

50% at 40

o

C

Calibration Default Settings (CAL dEF)

Conductivity

Calibration factor

Reset to 1.000

Factory Default Settings (FCt dEF)

Conductivity

Calibration factor

Reset to 1.000

Cell constant

1.00

Cell type

2-cell

Measuring range

0 to 2000 µS/cm

Line resistance

0.000

Temperature

Unit of measurement

o

C

Temperature compensation

ATC On

Temperature coefficient

2.1% (Both process & calibration temp.
coefficients)

Temperature coefficient type

Linear

MTC (ATC Off)

25

o

C (Both process & calibration

temperature)

Controller Output

Controller type

Limit control

Relay status

De-energized

Relay 1 (SP1)

Set point value

100 µS

Relay function

Low