2.7 Operating modes

3 operating
modes

The instrument has 3 operating modes:

- normal operation

- timed operation

- event operation

The following settings can, among others, be made for each of the three
operating modes:

- stored value

- storage cycle

Stored value

The stored value setting determines what is stored: average, minimum,
maximum or momentary value between two storage cycles, or the peak value
(envelope). The setting “peak value” means that the maximum and minimum
values from the last storage cycle are saved.

Storage cycle

The storage cycle determines the interval between two stored values. The
diagram speed corresponds to the storage cycle, which means that with a
storage cycle of 5sec, for example, the stored value is entered in the diagram
every 5sec.

Normal
operation

Normal operation is active whenever event or timed operation is not active.

Timed
operation

For timed operation, a period of time can be defined (up to 24 hrs) within
which a specific stored value and a specific storage cycle are active.

Event operation

Event operation is active as long as its control signal (

v Chapter 4.2.6

“Configuration - Recording”) is active. Event operation can be used, for
example, to shorten the storage cycle when an alarm is present.

Priority

The respective priorities of the operating modes are allocated as follows:

Active
operating mode

The active operating mode is shown in the diagram by the background color
for the diagram speed:

v Chapter 2.2 “Operating principle and graphic elements”

Operating mode

Priority

Normal
operation

low

Timed operation

average

Event operation

high

Operating mode

Color

Normal
operation

gray

Timed operation

turquoise

Event operation

orange