3 operation and visualization – West Control Solutions DataVU 5 User Manual

Page 52

Advertising
background image

3 Operation and visualization

52

Status
messages

Status messages from the CompactFlash manager are shown in the
corresponding action window. The following status messages are possible:

Status message

Description

CF CARD UPDATED

Directly before removing the CF card from the
instrument, it is necessary to call up Update CF
card
so that all measurement data up to the time
of removal are contained on the CF card. The data
not yet stored since the last automatic saving are
written to the CF card.

NO CF CARD

If there is no CF card in the instrument, the CF
card symbol flashes in the status line.

CF CARD FAULT

An error has occurred while writing to the CF card.
The CF card is faulty.
Remedy:
Insert new (FAT16-formatted) CF card.

CF CARD FULL

If the CF card is full, the CF card symbol flashes in
the status line. No more data will be written to the
CF card.
Remedy:
Insert a blank CF card before the measurement
data memory of the recorder is also full. If this is
not done, measurement data will be lost.

NO CONFIG. FILE ON CF

An attempt was made to read a configuration from
the CF card, but there is no suitable configuration
file on the CF card.

Advertising