4 configuration parameters – West Control Solutions DataVU 5 User Manual

Page 61

Advertising
background image

61

4 Configuration parameters

Channel
representation

Parameterization
➔ Diagram view

➔ Chan. represent.

Meas. small,
Scaling
Bar graph
Meas. large,
Off

This determines the
contents of the channel
line (header)

Channel indication

Parameterization
➔ Diagram view

➔ Channel indication

➔ Channel 1 — 6

Yes,
No

“Yes” means that the
selected channels are
shown in the header

Paper perforations

Parameterization
➔ Diagram view

➔ Paper perforations

No,
Yes

“Yes” means that paper
perforations are shown
within the graphic display,
on the left and right
margins of the screen.
Paper perforations can
only be activated and
shown if no event traces
have been selected (signal
type = analog input)

Counter/int. reset

Parameterization
➔ Reset counter/int.

➔ Channel 1 — 6

-999999999 to 0 to
+999999999

The start value for the
counter or integrator can
be entered here. The
current value is not saved.
Count changes are
documented in the event
list.
v Chapter 3.3

“Parameterization”

If the password query
is active (password “Reset
counter/int.” > 0), the start
values can only be set if
the password has been
entered correctly.

Parameter

Value/Selection

Description