3 Operation and visualization

54

-

This menu also includes the version number (e.g. 208.01.04). This is the

version number of the device software. It is important because some
functions are only available from a certain version number onwards.

- If the “Digital I/O option” is available (Yes), then the recorder also includes

an RS232/RS485 interface.

- If the Ethernet interface is available, then the field “Vers. No. Ethernet”

includes a number combination (e.g. 183.02.02). If this option is not
available, the field is filled with “-”.

Error

The following errors may occur:

Error

Description

none

no error in instrument

Data lost

A discharge of the battery or storage capacitor
occurred during the last lengthy power
interruption.

The clock was set to 01.01.00 00:00:00.

Remedy:
Reset the time (

v Chapter 4.2.2 “Configuration -

Instrument (device) data”) and use a new
CF card for data storage.

Battery empty

This message appears on instruments with a
lithium battery, when the time was reset after a
loss of data but the battery is still empty

Please return the instrument to the supplier for a
change of battery.

H

Data may be lost after separating the instrument from the supply:
after more than 10 years (on instruments with a lithium battery) or
after approx. 2 weeks (at ambient temperature 15 to 25°C) with a
storage capacitor.