Warning, Caution, Cartridge cleaning (figure 15) – Badger Meter Flow-Alert Reed Switch User Manual

Page 14

Advertising
background image

Page 14

Form #04-VAM-UM-00229 09/12

GPM

LPM

OIL

Mounting

Bracket

Mounting

Bracket

Screws

Strain

Relief

Switch

Mounting

Screw

Side

Bracket

Reed

Switch

Scale

Mounting

Screw

FIGURE 14 - REED SwItCH

REPlACEMENt

Cartridge Cleaning (Figure 15)

1. Disconnect the meter cable.
2. Remove the meter from the

line. Remove excess piping

from meter.

WARNING

Warning - Before attempting to

remove the meter from the line,

check the system to confirm that

line pressure has been reduced

to zero PSI. Failure to follow these

instructions could result in serious

personal injury or death and/or

damage to the equipment.

NotE: It is not necessary to remove the

aluminum housing from the meter to

remove it from the line.

3. Thoroughly wipe off the

entire meter surface using

mild detergent or isopropyl

alcohol.

CAUTION

Caution - Do not use aromatic

hydrocarbons, halogenated hydro-

carbons, ketones or ester based

fluids on polycarbonate lens. Failure

to follow these instructions could

result in damage to the meter.

4. Remove the inlet end

fitting, retaining spring, and

metering cone/spider plate

assembly from the cartridge.

5. Gently push the cartridge

assembly towards the outlet

port while holding magnetic

indicator assembly in place.

6. The cartridge internal parts

are secured with a retaining

ring. Remove the retaining

ring and the remaining

internal parts from the

cartridge.

NotE: If internal parts do not slide

freely from cartridge, use a wooden

dowel inserted into the outlet port of

the meter to push parts out.

7. Place all parts on a clean work

surface. Clean and inspect all

parts. Replace any that appear

worn or damaged.

8. Check inlet port O-ring

for damage and replace if

required.