To adjust Pressures beyond
(above or below) scale limits:
Step 1. Locate point at which the
brightly colored indicator line inter-
sects the vertical 100 PSIG pressure
line.
Step 2. Divide this reading by the
Pressure Correction Factor (f
1
) indi-
cated in the Conversion Chart.
To adjust for changes
in Temperature :
Step 1. Divide the 100 PSIG flow
rate reading by the Temperature
Correction Factor (f
2
).
To adjust for changes
in Specific Gravity:
Step 1. Establish the square root of
the new specific gravity.
Step 2. Divide the 100 PSIG flow
rate reading by the Specific Gravity
Correction Factor (f
3
).