Barometric reference (baro) module specifications – Crystal nVision Lab Reference User Manual

Page 26

Advertising
background image

Modules

24

nVision Operation Manual

BAROMETRIC REFERENCE (BARO) MODULE SPECIFICATIONS

Absolute pressure mode is achieved by using the BARO Reference Module to establish a datum. The PM and BARO sensor uncertainties were combined to
establish a new accuracy statement for readings taken in the absolute mode. It must be noted that the accuracy statement is valid for readings of 1 barA or
greater.

Accuracy

± 0.00725 PSI, ± 0.5 mbar

Accuracy specifications include all effects of linearity, hysteresis, repeatability, temperature, and stability within the specified operating temperature range
for one year.

Note:

Exposure to environmental extremes of temperature, shock, and/or vibration may warrant a more frequent recertification period.

Ranges, Resolutions, and Units

Range . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.153 to 15.954 PSIA, 700 to 1100 mbarA

Units and Resolution

PSI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.001

inHg . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.001

mmHg . . . . . . . . . . . . . . . . . . . . . . . . . . 0.01

mbar . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.1

Logging Interval

Fastest Logging Interval . . . . . . . . . . 10 readings per second

Pressure Connection

Cylindrical sensor fitting of 5.8mm OD. A flexible 4.8 mm [3/16"] ID tube is recommended to connection for both nVision chassis forms.

Mounting

BARO Module is secured using 3/8” 4-40 plastic screw.

!

WARNING: Plastic non-conductive screw must be used to comply with hazardous locations requirements.

!

CAUTION: Direct contact with barometric sensor may cause permanent damage. Direct sunlight on exposed BARO sensor may affect readings slightly.