BAROMETRIC REFERENCE (BARO) MODULE SPECIFICATIONS

Absolute pressure mode is achieved by using the BARO Reference Module to establish a datum. The PM and BARO sensor uncertainties were combined to
establish a new accuracy statement for readings taken in the absolute mode. It must be noted that the accuracy statement is valid for readings of 1 barA or
greater.

Accuracy

± 0.00725 PSI, ± 0.5 mbar

Accuracy specifications include all effects of linearity, hysteresis, repeatability, temperature, and stability within the specified operating temperature range
for one year.

Note:

Exposure to environmental extremes of temperature, shock, and/or vibration may warrant a more frequent recertification period.

Ranges, Resolutions, and Units

Range . . . . . . . . . . . . . . . . . . . . . . . . . . . 10.153 to 15.954 PSIA, 700 to 1100 mbarA

Units and Resolution

PSI . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.001

inHg . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.001

mmHg . . . . . . . . . . . . . . . . . . . . . . . . . . 0.01

mbar . . . . . . . . . . . . . . . . . . . . . . . . . . . . 0.1

Logging Interval

Fastest Logging Interval . . . . . . . . . . 10 readings per second

Pressure Connection

Cylindrical sensor fitting of 5.8mm OD. A flexible 4.8 mm [3/16"] ID tube is recommended to connection for both nVision chassis forms.

Mounting

BARO Module is secured using 3/8” 4-40 plastic screw.

!

WARNING: Plastic non-conductive screw must be used to comply with hazardous locations requirements.

!

CAUTION: Direct contact with barometric sensor may cause permanent damage. Direct sunlight on exposed BARO sensor may affect readings slightly.