Functions

4

nVision Operation Manual

Zero

To zero the nVision:

Press the

(

zero

)

button for at least 1 second while vented to atmosphere until the dashed lines (- - - - -) appear.

To clear the zero value:

Hold the

(

zero

)

button for 5 seconds until the display changes from (- - - - -) to the zeroed value, then to (- - -).

You can adjust the

Zero Limit

at which the

(

zero

)

button will display “- -HI- -” in

CrystalControl.

You can also disable the

(

zero

)

button

entirely, by setting the zero value to a negative number less than -15 PSI.

Note:

If you attempt to zero the gauge with more pressure applied than the Zero Limit set in with CrystalControl, the command will be ignored, and “- -HI- -”

will display.

Note:

You can never zero the BARO sensor.

!

WARNING: This gauge can display zero pressure when connected to a source of pressure! Do not rely on the display indication before disconnecting—

it may not be indicating true pressure. Never disconnect pressure instrumentation without first relieving system pressure!