Logging Interval

Fastest Logging Interval . . . . . . . . . . 5 readings per second.

Note:

Although the nVision logging interval may be set to a faster rate, the RTD100 module will update the reading at 5 times per second.

ATEX and IECEx Scheme Entity Parameters

The RTD100 module has these specific input entity parameters:

Ui = 0

Uo = 9.73 V

li = 0

lo = 1.6642 A

Pi = 0

Po = 1.1 W

Co = 0.5 uF

Lo = 12 uH

*

*

Total cable inductance between all modules

RESISTANCE TEMPERATURE DETECTORS (RTD)

Resistance Temperature Detectors (RTDs) are temperature sensors that contain a resistor that utilize the predictable change in electrical resistance of particu-
lar materials over temperature. Platinum elements have been used for many years in laboratories and industrial processes, and have a reputation for range,
linearity, repeatability, and stability. The selection strengths of RTDs, or sometimes called PRT (platinum resistance thermometer) are their wide temperature
range (approximately -200 to 850ºC), accuracy (better than thermocouples), good interchangeability between similar sensors, and long-term stability.

Callendar-Van Dusen Equation

The relationship between temperature and resistance is given by the Callendar-Van Dusen equation.

RT = R0 [1 + AT + BT

2

+ CT

3

(T-100)] for (-200°C < T < 0°C)

RT = R0 [1 + AT + BT

2

] for (0°C ≤ T ≤ °C of Upper Temperature Range listed below)

Where: RT = the resistance at temperature, T; R0 = the resistance at 0°C; and the constants A, B, and C dependent upon RTD selected (TCR).

nVision (TCR)

Temp. Coefficient of

Resistance

Temperature

Range

Base

Resistance

TCR (Ω/Ω/ºC)

Sensitivity
(avg. Ω/ºC,

0 to 100 ºC)

A (°C-1)

B (°C-2)

C (°C-4)

Pt100 (385) Euro

-200 to 850ºC

(-328 to 1562ºF)

100Ω at 0ºC

0.00385

0.385

3.9083 x 10

-3

-5.7750 x 10

-7

-4.183 x 10

-12

Pt100 (3926) US

-259 to 1235ºC

(-434 to 2255ºF)

100Ω at 0ºC

0.003926

0.3926

3.9848 x 10

-3

-5.87 x 10

-7

-4.0 x 10

-12

Pt100 (3911)

-259 to 630ºC

(-434 to 1166ºF)

100Ω at 0ºC

0.003911

0.3911

3.9692 x 10

-3

-5.8495 x 10

-7

-4.2325 x 10

-12