Crystal nVision Lab Reference User Manual

Page 31

Advertising
background image

Modules

29

nVision Operation Manual

Logging Interval

Fastest Logging Interval. . . . . . . 6 readings per second.

Note:

Although nVision logging interval may be set to a faster rate, the MA20 module will update the reading at 6 times per second.

ATEX and IECEx Scheme Entity Parameters

The MA20 Module has these specific input entity parameters:

Ui = 28 V

Uo = 6.6 V

li = 93.3 mA

lo = 4.45 mA

Pi = 653.3 mW

Po = 7.34 mW

Ci = 0.36 uF

Co = 0.5 uF

**

Li = 39.1 uH

Lo = 12 uH

*

*

Total cable inductance between all modules

**

Dependent on the supply to the terminals but shall not be greater than 0.5 uF

Current (mA) Input

Accuracy

±(0.015% of reading + 0.002mA)

Accuracy specifications include all effects of linearity, hysteresis, repeatability, temperature, and stability within the specified operating temperature range
for one year.

Note:

Exposure to environmental extremes of temperature, shock, and/or vibration may warrant a more frequent recertification period.

Ranges, Resolutions, and Units

Range . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 55mA (MA20+)
Range . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 25mA (MA20)

Max Allowable Current . . . . . . . . . . . 93.3mA

Note:

Inputs protected by resettable fuse

Resolution . . . . . . . . . . . . . . . . . . . . . . . 0.001mA or 0.01%

Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . mA, % 4-20, %10-50

mA Input

Input Resistance . . . . . . . . . . . . . . . . . < 17.2 Ω

Voltage Burden @ 50mA . . . . . . . . . < 0.86 V
Voltage Burden @ 20mA . . . . . . . . . < 0.35 V

HART mA Input

HART Resistance . . . . . . . . . . . . . . . . . 250 Ω