Voltage (V) Input

Accuracy

±(0.015% of reading + 0.002VDC)

Accuracy specifications include all effects of linearity, hysteresis, repeatability, temperature, and stability within the specified operating temperature range
for one year.

Note:

Exposure to environmental extremes of temperature, shock, and/or vibration may warrant a more frequent recertification period.

Ranges, Resolutions, and Units

Range . . . . . . . . . . . . . . . . . . . . . . . . . . . 0 to 28VDC

Max Allowable Voltage . . . . . . . . . . . 30VDC

Resolution . . . . . . . . . . . . . . . . . . . . . . . 0.001VDC

Units . . . . . . . . . . . . . . . . . . . . . . . . . . . . VDC

Switch Test

Switch Detection

Switch States . . . . . . . . . . . . . . . . . . . . Dry Contact

Closed State Resistance . . . . . . . . . . < 10 Ω

Open State Resistance . . . . . . . . . . . > 10 MΩ

TEMPERATURE (RTD100) MODULE INSTRUCTIONS

Your nVision has the ability to measure temperature very accurately if populated with an RTD100 module. With this system you may connect your resistance
temperature detector (platinum RTD) or platinum resistance thermometer (PRT) to the nVision using the provided IP67 rated connector system (P/N: 3953).
Once the sensing element is connected, you may display the temperature reading in your desired unit. The nVision can also measure electrical resistance (Ω)
to help in troubleshooting your resistance based sensing element.

Temperature Coefficient of Resistance (TCR)

Your nVision comes pre-loaded with several common RTD sensing element TCR values with the appropriate Callendar-Van Dusen Coefficients to convert your
resistance measurement to the appropriate temperature measurement. The available 100Ω platinum RTD TCRs are:

Pt100 (385) Euro

Pt100 (3911) US

Pt100 (3926)

Simply select the desired TCR or coefficient values in CrystalControl for use on the nVision. You may also view the Summary page to confirm

you have selected the correct setting for your sensing element.