Crystal nVision Lab Reference User Manual

Page 50

Advertising
background image

Support

48

nVision Operation Manual

Error 1 displayed

Reason

The nVision checks the integrity of internal calibration coefficients every time it’s turned on. If any coefficients have been corrupted in any way, “Error 1” is
displayed.

Solution

Contact factory for instructions on how to restore the memory to the original factory settings.

Error 2 displayed

Reason

The nVision has tried to display a number too large for the display (i.e., more than 6 digits). May be due to an electrical malfunction or numerical error.

Solution

Contact factory for further instructions.

Error 5 or Error 6 displayed

Reason

The nVision pressure module (PM) is exhibiting out of normal operating condition behavior.

Solution

Contact factory for module replacement.

Error 7 displayed

Reason:

The nVision has lost communication with one or both of your modules.

Solution

It may be possible to recover from this condition by removing and reconnecting your modules using the Module Installation Instructions in this manual. This
may happen if you change modules but do not remove and replace the 4AA, USB, or AC Power to the unit to force a complete reset of the product. Once this
is done the module should be recognized properly. If unsuccessful in resolving the issue, contact factory for module replacement.

Date and Time are incorrect

Reason

The nVision has its date and time synchronised to computers located in our factory.

Solution

Connect to CrystalControl on a computer with the correct date and time settings you desire. During the connection process, the nVision

will be synchronized to the local time.