Setup mode – Sper Scientific 860031 pH - mV Bench-Top Meter User Manual
Page 27

27
SETUP MODE
on the lower display.
3. Press to enter P8.1. The year flashes in the lower
left corner of the LCD display. (The year is the last two
digits only; for example, 1999 would be 99).
Symbol:
Y-M-D
H:M:S
Definition:
Yr.-Mo.-Day
Hr.-Min.-Sec.
Range:
99-12-31
23-59-59
4. Press
to step through the following “P’s.” All are
two digits.
P8.1 = Year
P8.2 = Month
P8.3 = Day
P8.4 = Hour
P8.5 = Minute
P8.6 = Seconds
5. Press ▲ and ▼ to adjust values up or down,
respectively.
6. Press ESC to return to P8.0.
7. Press ESC to return to Normal Mode.
Reset
This procedure will reset the meter to factory default
settings. Memory locations are not reset after this
procedure.
1. Press MODE continuously until you reach the mode that
you want to reset. When resetting pH/mV, only pH and
mV will revert to the default values. The COND/TDS/
SALT parameters will not be reset unless you select the