Salinity calibration – Sper Scientific 860031 pH - mV Bench-Top Meter User Manual
Page 37
Advertising

37
The procedure for salinity calibration is almost the same as
the procedure for conductivity calibration. Differences are
as follows:
1. In Measurement Mode, press MODE to select the mode
as SALT and press CAL to enter Calibration Mode.
2. There are two measuring ranges for salinity: 0 to 11.38
ppt and 0 to 80.0 ppt. Please select a NaCl standard
that is near the sample value you are measuring.
SALINITY CALIBRATION