2. Press MODE to select pH Mode.

3. To set the temperature, press

for more than 1

second. “CAL” will flash on the LCD.

4. Press or to change the temperature value. Press

to save and return to Normal Mode.

Conductivity Probe
The temperature sensor is built into the conductivity probe.
Follow the manual temperature setting procedures in P5.1
and P5.3 (pages 23-25) to set the temperature.

Note...
There is no need to take temperature compensation into
consideration when measuring ORP.

Auto and Manual Range


Press while in Normal Mode to select automatic or
manual range function.

Mode

CON

TDS

SALT

Auto

Full range

Full range

Full range

Range 1

0 ~ 19.99 uS 0 ~ 19.99*f ppm

Range 2

0 ~ 199.9 uS 0 ~ 199.9*f ppm

Range 3

0 ~ 1999 uS 0 ~ 1999*f ppm

Range 4 0 ~ 19.99 mS 0 ~ 19.99*f ppt

0 ~ 11.38 ppt

Range 5

0 ~ 199.9 mS 0 ~ 199.9*f ppt

0 ~ 80.0 ppt

Note…
“f” stands for TDS conversion factor.

Normally, the meter will automatically select a range while
taking readings. To select a specific range (or