Ph probe calibration – Sper Scientific 860031 pH - mV Bench-Top Meter User Manual

Page 30

Advertising
background image

30

PH PROBE CALIBRATION

7. If NIST is selected, the lower display indicates the value

of the solution at the current temperature.

Note...
If this secondary value continues to fluctuate, check the
buffer or probe. (Refer to TROUBLESHOOTING page 55).
If CUST is selected, the lower middle display indicates the
default, 2.00. Press HLD to select the buffer range needed.
Press or to adjust the lower middle display to coincide
with the main display reading.

8. When the measured pH value is stable and the Ready

function is enabled (Refer to P6.0 READY ICON page
25), “READY” will appear on the left side of the LCD.
Press

to confirm.

9. Change the buffer solution and repeat the previous

steps to achieve multiple point calibration. Clean the
probe in between each buffer.

10. Press ESC to return to Normal Mode.