Example 2:
Standard: 1.6 uS; Detected value: 2.1 uS
Adjustable range: ± 0.42 uS (2.1*20)
Although the measured value differs less than 20%, the 1.6
uS is still less than the 10% range limit (19.99*10%).
Therefore, the maximum input value is 2.00 uS.

CONDUCTIVITY PROBE CALIBRATION