Vaisala HMP228 User Manual

Page 28

Advertising
background image

HMP228
Operating Manual

M210282en-A

22

4.3.1

Two point calibration procedure

A two point calibration should be performed in stable conditions using
saturated salt solutions as references.

Disconnect the security lock jumper!

4.3.1.1

Using serial commands

•

Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

•

Insert the sensor head into the calibration hole of the LiCl chamber of

the humidity calibrator.

•

Wait at least 30 minutes.

•

Give command CRH <cr>, enter the first point value and press <cr>.

>CRH <cr>
RH : 11.9 Ref1 ? yy.y <cr>
Press any key when ready...

•

If you want to see how the sensor stabilizes to the humidity in the
calibrator, enter c <cr> instead of the first reference:

RH : 11.9 Ref1 ? c <cr>
RH : 11.5 Ref1 ? c <cr>
RH : 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...

•

Insert the sensor head into the calibration hole of the NaCl chamber of

the humidity calibrator.

•

Wait at least 30 minutes.

•

Press any key and enter the second point value; press <cr>.

RH : 75.5 Ref2 ? yy.y <cr>

•

The stabilization of the sensor can be monitored by entering c <cr>

instead of the reference value.

4.3.1.2

Using display/keypad commands

•

Leave the calibrator and the transmitter for at least 4 hours in the
same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

•

Select Cali in the main menu and then RH cal; select Not changed
and then two point calibration RH 2 point cal. Change the first point
reading with the arrow keys to correspond to the reference humidity
and press ENT; pressing an arrow once changes the reading by
0.05 %RH.