Vaisala HMP228 User Manual

Page 68

Advertising
background image

HMP228 Transmitters
Appendix 1: Serial commands

M210282en-A

62

MTIM

Setting the measurement integration time

MTIM nnn <cr>

nnn

=

number of cycles measured (4...255)

By lengthening the measurement integration time any stray changes in the out-
put can be filtered out: the transmitter calculates the average of a number of
measurement cycles defined by the user. The command can be given in two
ways:

>MTIM <cr>
Mtim : 4 ? 5 <cr>

or

>MTIM 5 <cr>
Mtim : 5

PRES

Setting the pressure for mixing ratio calculations

PRES pppp.pp <cr>

pppp.pp =

pressure (hPa)

The atmospheric pressure has an effect on mixing ratio. Therefore accurate
mixing ratio calculations can be achieved only when the ambient pressure is
taken into consideration.

When the command is given, the transmitter first gives the currently used
pressure; after this a new value can be entered or the old one acknowledged:

>PRES <cr>
Pressure : 1013.25 ? 1000.00 <cr>

When the currently used pressure is known, a new pressure can also be entered
directly:

>PRES 1010 <cr>
Pressure : 1010

NOTE

If the pressure setting is frequently adjusted, e.g. by
using an external barometer as a pressure input source,
the command XPRES is recommended.

Advertising