31
T = D0 +R x {D1 + R x [D2 + R x (D3 + R x D4)]}
where
D0
=
-243.5673014
D1
=
2.278542701
D2
=
0.002050681
D3
=
-6.15025E-06
D4
=
1.34949E-08
4.7
Analogue output channels
4.7.1
Setting the analogue outputs
The HMP228 transmitters can be ordered with the required current or voltage
outputs already selected. If the outputs need to be changed, move the jumpers
in connector X15 into positions as shown in Figure 4.7.1.2.
24V
Sensor connections
Current loop or
RS485 module
Ch1 Ch2
+ -
nc
Tx
GND
Rx
-
- +
+
X55
Spare
jumpers
Figure 4.7.1.1
Spare jumpers