HMP228 Transmitters
Appendix 1: Serial commands
M210282en-A
48
The output scales can also be given directly with the ASCL command.
ASCL aaa.a bbb.b ccc.c ddd.d <cr>
aaa.a
=
lower limit of channel 1
bbb.b
=
upper limit of channel 1
ccc.c
=
lower limit of channel 2
ddd.d
=
upper limit of channel 2
For example, water activity is scaled to 0...1.00 on channel 1 and temperature
to -40...+100
°
C on channel 2:
>ASCL 0 1 -40 100 <cr>
Ch1
(aw) lo
0.000
Ch1
(aw) hi
1.000
Ch2
(T ) lo
-40.000 'C
Ch2
(T ) hi
100.000 'C