Vaisala HMP228 User Manual

Page 56

Advertising
background image

HMP228 Transmitters
Appendix 1: Serial commands

M210282en-A

50

The command can also be divided into two commands, so the computer can be
used for other purposes while waiting for the sensor to stabilize to the higher
humidity.

>FCRH 1 <cr>
RH : 1.90

Ref1 ?

11.3 <cr>

FCRH 2 <cr> ...
RH : 76.30

Ref2 ?

74.9 <cr>

If the stabilization of the sensor to the humidity in the calibrator needs to be
monitored, the measurement output can be repeated by giving command c<cr>
at Ref1 and Ref2:

>FCRH <cr>
RH : 12.00

Ref1 ? c <cr>

RH : 11.70

Ref1 ? c <cr>

RH : 11.50

Ref1 ? 11.3 <cr>

Press any key when ready ...
RH : 76.00

Ref2 ? 75.5 <cr>

CT

Temperature calibration

Disconnect the security lock jumper!

CT <cr>

Using this command the transmitters can be calibrated against an accurate ref-
erence, such as a Pt 100 simulator. A two-point calibration is performed as
follows:

>CT <cr>
T : 0.80

Ref1 ?

0.0 <cr>

Press any key when ready ...
T : 56.20

Ref2 ?

55.0 <cr>

In one-point offset correction, the Ref2 prompt is acknowledged with <cr>:

>CT <cr>
T : 0.80

Ref1 ?

0.0 <cr>

Press any key when ready ...
T : 75.50

Ref2 ?

<cr>

If the stabilization of the sensor to the temperature of the calibrator or the ref-
erence needs to be monitored, the measurement output can be repeated by giv-
ing command c<cr> at Ref1 and Ref2:

>CT <cr>
T : 0.80

Ref1 ?

c <cr>

T : 0.40

Ref1 ?

0.00 <cr>

Press any key when ready ...
T : 56.20

Ref2 ?

55.0 <cr>