Vaisala HMP228 User Manual

Page 31

Advertising
background image

HMP228

M210282en-A

Operating Manual

25

4.4.1

Calibration procedure after sensor change

Humidity calibration should be performed in stable conditions using saturated
salt solutions as a reference.

Disconnect the security lock jumper!

4.4.1.1

Using serial commands

•

Leave the calibrator and the transmitter for at least 4 hours in the

same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

•

Insert the sensor head into the calibration hole of the LiCl chamber of

the humidity calibrator.

•

Wait for 30 minutes.

•

Give command FCRH <cr>, enter the first point value and press <cr>:

>FCRH <cr>
RH: 11.9 Ref1 ? yy.y <cr>
Press any key when ready...

•

The stabilization of the sensor to the reference humidity can be

monitored by giving c <cr>:

RH: 11.9 Ref1 ? c <cr>
RH: 11.5 Ref1 ? c <cr>
RH: 11.5 Ref1 ? 11.3 <cr>
Press any key when ready...

•

Insert the sensor head into the calibration hole of the NaCl chamber of

the humidity calibrator.

•

Wait for 30 minutes.

•

Press any key, enter the second point value and press <cr>:

RH: 75.5 Ref2 ? yy.y <cr>

•

The stabilization of the sensor can be monitored by entering c <cr>

instead of the reference value.

4.4.1.2

Using display/keypad commands

•

Leave the calibrator and the transmitter for at least 4 hours in the

same space so that their temperatures have time to equalize. Remove
the filter cap on the transmitter.

•

Insert the sensor head into the calibration hole of the LiCl chamber of

the humidity calibrator.

•

Wait for 30 minutes.