HMP228 Transmitters
Appendix 1: Serial commands
M210282en-A
64
TIME
Setting the time
TIME <cr>
For example, to enter a new time:
>TIME <cr>
Current time is 01:35:54
Enter new time (hh:mm:ss) : 13:25:56 <cr>
When the current time is asked, the new time is passed with <cr>.
VERS
Name and version of the programme
VERS <cr>
For example:
>VERS <cr>
HMP 228 / x.yy
where x.yy is the programme version.
?
Outputting the transmitter settings
? <cr>
For example:
>? <cr>
HMP 228 / 1.01
CPU serial nr : 0
Keyboard type : 0
Address
: 7
Output units
: metric
Baud P D S
: 4800 E 7 1 FDX
Serial mode
: STOP
Output intrv. : 0 min
Mtim
: 32
Pressure
: 1013.25
Analog outputs
Ch1 0.00 ...
10.00 V
Ch2 0.00 ...
10.00 V
Ch1 ( aw )
lo
0.000
Ch1 ( aw )
hi
1.000
Ch2 ( T )
lo
-20.000'C
Ch2 ( T )
hi
180.000'C
Transducer
:
PRB serial nr : 0
Calibr. date
: 0
??
Outputting the transmitter settings also in POLL mode