Enter the following:

a

Disconnect After Scan: To maintain a persistent con-

nection to the device, choose False.

b

Connection Type: If the transmitter has a LAN or

WLAN module, select Network. If you use an external

Digi device, choose COM Port.

c

IP Address: See your administrator or check the De-

vice box.

d

Sample Rate: Choose 90 (recommended, but you can

modify the rate if you need more or less recorded da-

ta).

e

Connection Timeout: To ensure continuous monitor-

ing, do not change. Contact your network administra-

tor if required.

4

To add several 300 Series Transmitters, HMT140 Series

Transmitters or a combination of Vaisala Veriteq devices

and 300 Series Transmitter devices, select Upload

Definitions File then enter or browse to the correct file. To

create a Definitions File, see the FAQ on page 145.

Swapping Devices

If you need to send a device for maintenance (such as

recalibration), you will need to swap it for another device to

ensure continuous monitoring and alarming.

Note:

If a device is removed without being swapped,

this will trigger a communication alarm.

When a device is swapped, the change is noted on the

Location History report (the report shows the device serial

number for a reporting period). If, during the reporting period,

the device was swapped, this event is listed in the report

summary.

To swap a device:

1

Ensure the new device has the same settings as the

device to be swapped (see

Options | System

Configuration | Locations Manager

, or refer to your

Vaisala product User Guide).