About Alarms

viewLinc 4.1 Administrator Guide

83

Event Log Validation Alarms

An Event Log Validation Alarm indicates when changes have

been made to the event log from outside the viewLinc

system. Each entry in the event log has its own CRC which is

verified - if any changes have been made, an alarm is

automatically created, rendering data non-validatable. It

checks the CRC of event files, checks the CRC of event

records in the files, checks for missing records, and checks

for missing files. If any are discovered, an alarm is raised

and must be acknowledged.

Device Configuration Alarms

If you receive a Configuration Alarm, this indicates that your

device has stopped recording data history or was configured

incorrectly. This could be the result of being set to stop when

full, a delayed start, or the device could have an internal

error. This alarm can also be triggered if a channel has been

disabled in the device using vLog. To correct this issue, you

can verify or modify the Vaisala Veriteq device settings using

vLog. If the problem persists, contact your Vaisala technical

support representative. The default delay for configuration

alarms is 60 seconds. However this delay is bypassed for

certain alarm types such as Missing Channel alarms,

Configuration changes and Probe changes.

Device Validation Alarms

If you use a VL-type data logger, you’ll receive a Validation

alarm if the validation memory in the device is corrupted or

has been modified. Contact your Vaisala technical support

representative.

Device Calibration Alarms

If you use a VL-type device, Calibration Alarms send you

intermittent notification when your device is due for

calibration. You receive notifications at the following intervals:

3 months and 1 month before the calibration due date, then

again on the data logger’s scheduled recal date. This alarm

will remain active until the logger has been recalibrated.