Vaisala viewLinc-4.1-M211518 User Manual

Page 54

Advertising
background image

Swapping Devices

viewLinc 4.1 Administrator Guide

43

Note:

Each Vaisala Veriteq device must be reviewed

separately using vLog, and the COM port to

which they are connected must be selected.

COM port/IP address

.

Device is on the same COM port

(attached to the same cable). If you are swapping a 300

Series Transmitter connected as a network device,

ensure the new transmitter has the same IP address.

Name

.

Device is not retired or deactivated.

Status

.

Device is not in active alarm mode.

Sample Interval

.

Devices have the same sample

interval.

Channels

.

Devices have the same channels enabled

and be using the same units of measure for each

matching channel.

Audit Trail

.

Vaisala Veriteq devices are not linked to a

vLog Audit Trail.

2

From

Options | Locations

, navigate the system tree to

find the device you want to swap. Right click and select

Pause Device Alarming

.

3

In the Pause Device Alarming screen, you can add a

comment about why you are swapping devices and

specify the expected duration for the pause. This will

ensure no communication or threshold alarms are

triggered when disconnecting the device.

Note:

When a device is paused any active alarms at

the time of the pause are turned off. Any event

logs and emails sent out will not indicate they

were turned off due to a pause.

4

Click

OK

.

Advertising