Lukas P 600 OE User Manual

Page 27

Advertising
background image

27

Fault

Check

Cause

Solution

P600OE stops
during operation

Battery or mains
unit positioned
correctly in the
connection slot
and locked in
position

Battery or mains unit
not inserted properly

Remove the battery
or mains unit from the
pump and re-insert.

Mains unit temperature
too high as a result
of continuous use
and high ambient
temperature

Allow mains unit to cool
down (see operating
instructions of mains
unit)

Electric motor not
providing full power

Check connection
of mains unit

Mains unit not correctly
connected

Re-insert the mains unit

Cable losses in
extension cable or
cable drums too great
(electrical resistance)

Use a different suitable
extension cable or cable
reel or connect directly
to power supply

Mains unit fault

Use a battery or
different mains unit.

Check battery
charge status

Battery almost fl at

Charge battery

Battery defective

Replace battery

Fault in power supply to
the mains unit

Troubleshooting the
power supply

Use a different power
supply

P600OE defective

Shut down immediately
and have repaired by
authorised service
facility, authorised
dealer or directly by
LUKAS