Correct use – IKA HBC 10 baisc User Manual

Page 25

Advertising
background image

25

• After a power failure during operation, the device may

start automatically (depending on operating mode).

• Transport the device with care.
• Do not transport or empty the bath while it is still hot.

This may result in accidents, especially scalding injuries.

Fluids

Only use fluids, which fulfill the re-
quirements for safety, health and
device compatibility. Be aware of

the chemical hazards that may be associated with the
bath fluid used. Observe all safety warnings for the fluids.

• Depending on the bath fluid used and the type of opera-

tion, toxic or flammable vapors can arise. Ensure suit-
able extraction.

• Do not use any fluid which may cause dangerous reac-

tions during processing.

• Only use recommended bath fluid. Only use non-acid

and non corroding fluid.

Only process and heat up any fluid
that has a flash point higher than
the adjusted safe temperature limit

that has been set. The safe temperature limit must al-
ways be set to at least 25 ºC lower than the flash point
of the fluid used. Examine regularly the function of the
safety temperature limiter.

Never operate the device without
sufficient heat carrier fluid! You
should also be careful to ensure

that the minimum clearances and immersion depths in
the fluid are observed. Check the fluid level detection at
a regular basis (see the chapter "Filling and draining").

• Continuous monitoring of the bath and the filling level of

the bath fluid is required, especially at high temperatures.

• For optimum temperature stability, the fluids viscosity

should be 50 mm

2

/s or less at its lowest operating tem-

perature. This permits good fluid circulation and mini-
mizes heating from the pump.

If water is used at higher tempera-
ture, there is heavy loss of fluid due
to the condensation of evapora-

tion steam.

• Untreated tap water is not recommended. It is recom-

mended to use distilled water or high purity water (ion
exchangers) and add 0,1 g soda (sodium carbonate
Na

2

CO

3

) /liter, to reduce corrosive properties.

Risk of burning caused by vapor or
hot water at the outlet of the cool-
ing coil.

Do not use the cooling coil with
water at bath temperatures > 95°
C.

For bath temperatures > 60°C make
sure that the flow rate through the
cooling coil is high enough.

Don't use following fluids:

- Untreated tap water
- Concentration of acids or bases
- Solutions with halides: chlorides, fluorides, bromides,

iodides or sulfur

- Bleach (Sodium Hypochlorite)
- Solution with chromates or chromium salts
- Glycerine
- Ferrous water.

When changing the bath fluid
from water to a heat transfer fluid
for temperatures above 100 °C, re-

move the remaining water from the complete system
(including hoses and external devices). When doing this,
also open the stopper and union nuts caps of the pump
outputs and inputs and blow compressed air through
all the pump outputs and inputs! Beware of the risk of
burning due to delay in boiling!

CAUTION

DANGER

WARNING

DANGER

CAUTION

DANGER

DANGER

CAUTION

WARNING

Correct use

• Use

Use IC (Immersion Circulator) and HBC (Heating Bath

Circulator)

- For heating and circulating fluids

Intended Use: Tabletop Device

• Range of use (indoor use only)

- Laboratories

- Schools

- Pharmacies

- Universities

This device is suitable for use in all areas except:
- Residential areas
- Areas that are connected directly to a low-voltage supply

network that also supplies residential areas.

The safety of the user cannot be guaranteed:
- If the device is operated with accessories that are not

supplied or recommended by the IKA

®

.

- If the device is operated improperly or in contrary to the

IKA

®

specifications.

- If the device or the printed circuit board are modified by

third parties.

Advertising
This manual is related to the following products: