Error codes

Any malfunctions during operation will be identified by an error message on the display.

Proceed as follows in such cases:

F

Switch off device using the main switch at the back of the device

F

Carry out corrective measures

F

Restart device

Error code

Effect

Cause

Solution

Err 01

Pump off
Heating off

No External Pt 100 temperature
sensor

- Check this sensor

Err 02

Pump off
Heating off

Motor Over current (rate
current)

- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked

Err 03

Pump off
Heating off

Motor Over current (Max
current)

- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked

Err 04

Pump off
Heating off

Motor Hall signal missing

- Reduce pump motor speed
- Use fluid with lower viscosity
- Check if the pump impeller is blocked

Err 05

Pump off
Heating off

Too high liquid level

- Check the liquid level and buoyage

Err 06

Pump off
Heating off

Too low liquid level

- Check the liquid level and buoyage

Err 07

Pump off
Heating off

Too high voltage

- Check the mains power

Err 08

Pump off
Heating off

Too low voltage

- Check the mains power

Err 09

Pump off
Heating off

Device internal temperature is
too high

- Check the ambient temperature and let the device

cool down

Err 10

Pump off
Heating off

PC communication failure

- Check communication cable

Err 11

Pump off
Heating off

Temperature difference
between control sensor and
safety sensor is too much

- Check safety temperature circuit and bath fluid

Err 12

Pump off
Heating off

Safety temperature alarm

- Check the bath temperature measurement

Err 13

Pump off
Heating off

Heater switched off by safety
circuit

- Check safety temperature set value, fluid level

Err 14

Pump off
Heating off

Fan error

- Check the fan and clean the grids at the rear side

If the actions described fails to resolve the fault or another error code is displayed then take one of the following steps:
- Contact the service department;
- Send the device for repair, including a short description of the fault.