Warning – IKA HBC 10 baisc User Manual

Page 27

Advertising
background image

27

Fig. 10

Fig. 11

Low Level

High Level

WARNING

• Fluids

No.

IKA

®

Designation

Operating temperature range

(°C)

Safety temperature

(°C)

Flash point

(°C)

0

EG19_N10.100 (H

2

0 + Glycole)

-10 ... 100

100

125

1

EG34_N20.100 (H

2

0 + Glycole)

-20 ... 100

100

125

2

EG40_N25.100 (H

2

0 + Glycole)

-25 ... 100

100

125

3

EG45_N30.100 (H

2

0 + Glycole)

-30 ... 100

100

125

4

SI_P20.275 (Silcone Oil)

20 ... 275

275

300

5

SI_N30.145 (Silicone Oil)

-30 ... 145

145

170

6

SY_N10.100 (Synthetic Oil)

-10 ... 100

100

125

7

Water*

5 ... 95

95

8

Customized

*

Note: Tap water may be unsuitable for operation because the calcium carbonate content may cause calcification.

High purity water (from icon exchangers) and distilled or bi-distilled water are unsuitable for operation due to corrosive
properties of these media. High purity water and distillates are suitable as a medium after adding 0,1 g soda (Na

2

CO

3

,

sodium carbonate) per liter of water.

Fig. 9

• Filling and draining

- Before filling the fluid into the bath, open the front

cover as indicated in following figure.

- Check and make sure that the drain valve is closed (the

stop position in clockwise direction, see Fig. 10).

Note: Please note information in chapter “Commis-

sioning”.

- Connect the mains plug and turn on the device with

mains switch (2).

- The low level warning message appear on the display.
- Open the lid (16) and fill fluid to the bath.

Note: Pay attention to the fluid

level information on the display:

- To drain the fluid from the bath, connect a hose to the

drain port and turn the drain valve in counter clockwise
direction with a straight screwdriver.

Drain valve

Drain port

Stop position