5 mode of operation – Eppendorf Multiporator User Manual
Page 32

74
Step 2 – Starting electroporation
–
Trigger the
electroporation
process by pressing
the START key.
–
"Charge" appears in
the third line of the
display. Charging
commences.
–
After the charging
procedure has
ended, discharging
occurs. This is
indicated by a flash
in the third line of
the display,
accompanied by a
double acoustic
signal.
After the experiment has
ended, a double acoustic
signal is emitted and the
initial information and
the set parameters
appear in the display.
The actual parameters
U
a
and
τ
a
appear in the
third line
of the display.
The time which has
elapsed since the end of
the experiment appears
in the bottom line in
minutes and seconds.
After 99 minutes or after
any key has been
pressed, this information
disappears from the
display.
U
a
: Actual measured
voltage used, in V
(volt).
τ
a
: Actual time constant
used, in ms
(milliseconds). Values
≤
0.8 are shown as
0.8 ms.
12.AUG.00
13:42
U 2000V
τ
5
ms
Charge
12.AUG.00
13:42
U 2000V
τ
5
ms
U 2003V
τ
5.1
ms
a
a
READY SINCE 1:36
5 Mode of operation
Multiporator_Text_en.fm Seite 74 Freitag, 9. Juni 2006 8:05 08