77

5.3.2 Procedure for cell fusion

Step 1 – Entering / modifying parameters

Press the SET
key to select the
parameters desired
(the parameter will
be underlined).

Use the arrow
keys to modify the
parameter in the
direction required.
According to the
parameters selected
for U'

~

and U"

~

, fusion

can be carried out
without alignment or
in combination with
one or both
alignment functions
as required.

Terminate the
entering procedure.
The parameters
are effective
immediately and
do not have to be
stored.

Example of a display of
electrofusion with activa-
ted alignments U'

~

/ U"

~

.

Example of the display
for carrying out elec-
trofusion with inactiva-
ted alignment U'

~

.

The relevant parameters
appear in the display in
larger letters and in bold
print.
U'

~

/ U"

~

:

Alternating voltage, in
V (volt).
Can be set to "0" and
in increments of 0.1 V
in the range 1.0–10 V.

t: Duration of alignment

0–95 s, in
s (seconds). Can be
set in increments of
5 seconds.

U : Voltage, in V (volt).

Can be set to "0" and
in increments of 1 V
between 5 V and
100 V, and in incre-
ments of 10 V be-
tween 100 V and
300 V.

t

: Duration of pulse,

in µs (microseconds).
Can be set in
increments of 5 µs
between 0 and
15 µs–300 µs.

n: Number of pulses

0–99.
Can be set in one-
pulse increments.

12.AUG.00

13:42

U'

~

5V

t

30s

U

30V

t

15µs

n2

U

"

~

5V

t

30s

12.AUG.00

13:42

U'

~

0V

t

30s

U

30V

t

15µs

n2

U

"

~

5V

t

30s

5 Mode of operation