77

5.3.2 Procedure for cell fusion

Step 1 – Entering / modifying parameters

–

Press the SET
key to select the
parameters desired
(the parameter will
be underlined).

–

Use the arrow
keys to modify the
parameter in the
direction required.
According to the
parameters selected
for U'

~

and U"

~

, fusion

can be carried out
without alignment or
in combination with
one or both
alignment functions
as required.

–

Terminate the
entering procedure.
The parameters
are effective
immediately and
do not have to be
stored.

Example of a display of
electrofusion with activa-
ted alignments U'

~

/ U"

~

.

Example of the display
for carrying out elec-
trofusion with inactiva-
ted alignment U'

~

.

The relevant parameters
appear in the display in
larger letters and in bold
print.
U'

~

/ U"

~

:

Alternating voltage, in
V (volt).
Can be set to "0" and
in increments of 0.1 V
in the range 1.0–10 V.

t: Duration of alignment

0–95 s, in
s (seconds). Can be
set in increments of
5 seconds.

U : Voltage, in V (volt).

Can be set to "0" and
in increments of 1 V
between 5 V and
100 V, and in incre-
ments of 10 V be-
tween 100 V and
300 V.

t

: Duration of pulse,

in µs (microseconds).
Can be set in
increments of 5 µs
between 0 and
15 µs–300 µs.

n: Number of pulses

0–99.
Can be set in one-
pulse increments.

12.AUG.00

13:42

U'

~

5V

t

30s

U

30V

t

15µs

n2

U

"

~

5V

t

30s

12.AUG.00

13:42

U'

~

0V

t

30s

U

30V

t

15µs

n2

U

"

~

5V

t

30s

5 Mode of operation