5.3 Mode for cell fusion (optional)
This mode is represented in the display by the symbol
.
Instructions for the usage of the Micro fusion chamber and the Helix
fusion chamber have been summarized in chapter 4.8 of the operating
instructions. A detailed description of the electrofusion of eukaryotic cells
can be found in the Basic Applications Manual for electrofusion, which is
included in the delivery package of the fusion module. This application
manual can also be downloaded from the Eppendorf Homepage
www.eppendorf.com.
5.3.1 Function: Cell alignment
The functions for alignment are indicated by the symbols U'
~
(before the
pulse, alignment) and U"
~
(after the pulse, post-alignment) in lines 2 and 4
of the display.
Step 1 – Entering / changing parameters
–
If necessary, switch
to the mode by
pressing the MODE
key.
–
Press the SET key
to select the
parameter desired
(the parameter will
be underlined).
–
Use the arrow keys
to modify the
parameter in the
direction required.
If a value in lines
U'
~
/ U"
~
/
U are
set to "0", the
function in question
is not carried out.
This enables U'
~
/
U"
~
to be triggered
individually or
consecutively
without the fusion
pulse being
activated.
Example of the display for
cell alignment U'
~
with-
out activation of pulse U
and alignment U"
~
.
The relevant parameters
appear in larger letters
and in bold print.
U'
~
/ U"
~
:
Alternating voltage, in
V (volts). Can be set
to "0" and in incre-
ments of 0.1 V be-
tween 1.0 V and 10 V.
t: Duration of alignment
0–95 s, in s (seconds).
Can be set in incre-
ments of 5 seconds.
U : Voltage,
(square-wave pulse)
unit V (volts).
Can be set to "0" and
in increments of 1 V
between 5 V and
100 V, and in
increments of 10 V
between 100 V and
300 V.
12.AUG.00
13:42
U'
~
5V
t
60s
U
0V
t
15s
n2
U
"
~
0V
t
30s
5 Mode of operation