8c. For “normally open” (NO) switching (code C switches), subtract 3 mm from
the measurement taken in step 7 and record the result. This distance will be
used to adjust the reed switch along the internal brass rod. The 3 mm is
subtracted to compensate for both the offset of the reed switch from the
actuation point (+3 mm), and the plug in the bottom of the stem (-6 mm). Refer
to drawings below.

Indention designates

top of float

For normally open (NO)

switching action, the center

point of reed switch is set 3 mm

above the desired actuation point.

ON

OFF

3 mm

9 mm

9 mm

Actuation

Point

Normally Open Switching for Code “C” Switches

Plug

Approx. 6 mm

Brass Rod