7000-50 & 7000-75 Series Operations Manual Rev A 1/6/2014

Page 12 of 71

Leveling the Unit

Place a level on the middle shelf of the incubator. Adjust the feet until the unit sits level

left to right and front to back. Even if the unit is level without adjustment, the leveling

feet should still be lowered to avoid the cabinet moving while opening and closing the

outer door & prevent a flat spot from forming on the casters.

Connecting the Drain Line

The chamber drain connection is located in the bottom middle of the back of the

chamber. A 1/2” tube fitting and tubing are supplied in the unit parts kit. Slide the

tubing into the drain connection. Pull on the tubing after installation to make sure it is

tight. Route the drain tubing to a local floor drain.

If a local floor drain is not available, a variety of accessories are available through

CARON customer service. These accessories can also be viewed at

www.caronproducts.com

.

For non-humidified chambers operated above 15°C, a drain is not needed. Install the

1/2” plug into the drain connection.

When using a pressurized water source, failure to connect

the unit to a drain could result in facility flooding.

The drain line relies on gravity to remove water from the

chamber. The drain line must remain below the chamber

to drain properly. Kinks or elevations in the drain line

above the cabinet drain will not allow the chamber to drain.

Do not have multiple loaded shelves out

simultaneously or the chamber may tip.