7000-50 & 7000-75 Series Operations Manual Rev A 1/6/2014

Page 27 of 71

SECTION 6 – ACCESSORY OPERATION

Operation of the Defrost System (DEFR303)

A Defrost System is used to maintain operational temperature below 5ºC. When the

chamber temperature is below 4.5ºC, the defrost system is enabled automatically.

When the chamber temperature is above 4.5ºC (except during a defrost cycle), the

defrost system is disabled automatically.

When the defrost system is enabled (below 4.5ºC), a defrost cycle will occur once

every 12 hours and lasts for about 20 minutes. During the defrost cycle, the chamber

temperature will increase several degrees above set point to melt any ice that has

formed on the evaporator. This temperature ‘spike’ is normal and assures long-term

operation.

During a defrost cycle, the

(Info) icon will appear on the status bar.

Press the

(Close Window) button to return to the main screen.