7000-50 & 7000-75 Series Operations Manual Rev A 1/6/2014

Page 16 of 71

Incoming compressed air must not exceed 100 psig.

CARON recommends installing a pressure gauge, filter, and shutoff prior to the dryer

to monitor incoming air pressure. The air supply (either house air or other compressed

air source) must be 90 to 100psig at the air supply inlet. An oil-less compressor may

be used. A Sound Suppression Kit maybe used to reduce the dryer noise (contact

CARON service for details).

1. Attach Dryer Package to exterior chamber with four mounting screws.

2. Connect dryer package power cord and 3/8” orange tubing into the dryer panel

on the chamber

Mounting

Screws

Mounting

Screws

Power Cord

Connection

Tube

Connection