7000-50 & 7000-75 Series Operations Manual Rev A 1/6/2014

Page 31 of 71

to the metal pen arm using a special "U" clip tab. For ease of replacement, it is

suggested that the two screws that hold the pen arm be loosened and the pen

cartridge and metal pen arm be removed as an assembly. Unsnap the plastic "U" clip

tab of the pen cartridge from the metal pen arm, remove and discard the old pen

cartridge. Replace the new cartridge by opening the hinge and snapping it securely

around the metal pen arm. Refer to the image below:

Pen Arm Calibration:

To check and/or adjust the recording pen(s) calibration to the outer most temperature

graduation of the chart, press and hold the "change chart" button (#3) until the pen

begins to move off of the chart. Once the pen(s) has moved off of the chart, again

press and hold the "change chart" button (#3) until the pen begins to move back onto

the chart. The pen should briefly stop at the outer most temperature graduation of the

chart before continuing onto the chart to begin recording. If the pen does not stop

exactly at this location on the chart, it can be adjusted or "calibrated" by using the left

(#1) or right (#2) arrow buttons.

When the pen moves back onto the chart and briefly stops, you will have

approximately five seconds in which to adjust the pen’s position using the left and right

arrow buttons of Figure 3.

On multiple pen recorders, each pen will move (one at-a-time) onto the chart briefly

stopping at the outer most temperature graduation of the chart at which time the pen’s

position can be adjusted by using the left (#1) or right (#2) arrow buttons. When the

time to adjust the position of the first pen has expired, the second pen will move onto

the chart briefly stopping at the outer most temperature graduation of the chart at

which time the second pen’s position may be adjusted.

Each time the chart paper or fiber tip pen cartridge is changed, you should make sure

that each pen stops at the outer most temperature graduation of the chart paper.

Otherwise, this pen offset will cause the unit to record an incorrect temperature on the

chart.