SECTION 8 – ALARMS

Alarm System Overview

The chamber control system is equipped with an alarm system that constantly monitors

temperature, and humidity (on humidified models) to ensure the user is notified if the

cabinet goes into an alarm condition. Notification occurs via an alarm pop-up window

and a buzzer. Each alarm condition has been factory programmed to minimize

nuisance alarms while maximizing warning time. There is a 2 hour time delay at start-

up and set point changes. To avoid nuisance alarms after a routine door opening, an

alarm condition must be present for 15 minutes before the operator is alerted. If the

optional remote alarm contacts are present, in an alarm condition, the dry contacts will

change state.

The following alarm messages could be displayed:

Chamber temperature is higher than set-point temperature

Chamber temperature is lower than set-point temperature

Chamber humidity is higher than set-point humidity

Chamber humidity is lower than set-point humidity

Temperature sensor error

In the event an alarm occurs, the alarm indicator will appear on the status bar and an

audible alarm will occur. The flashing (Alarm) icon will appear on the status bar and

the alarm pop-up window will automatically appear.