7000-50 & 7000-75 Series Operations Manual Rev A 1/6/2014

Page 42 of 71

Temperature calibration (example)

If the chamber temperature display reads 40.0

C and the calibrated independent

sensor shows 40.3

C, set the temperature offset value to 0.3

C. If the calibrated

independent sensor shows 39.6

C, then the entered offset should be negative. In this

example the required offset to temperature would be -0.4

C.

Calibrating the Humidity

If humidity calibration is needed, the following steps can be taken:

Locate the reference instrument’s humidity sensor in close proximity to the cabinet’s

geometric center. Be sure that the stabilization times described earlier have been

satisfied prior to performing this calibration.

A positive value will move the humidity ‘up’ and a negative value ‘down’. Press the

‘home’ button and verify the proper humidity is displayed.

Humidity calibration (example)

If the chamber temperature display reads 80% and the calibrated independent sensor

shows 83%, set the humidity offset value to 3.0%. If the calibrated independent sensor

shows 74

C, then the entered offset should be negative. In this example the required

offset to humidity would be -6.0%.

Calibrating Optional Chart Recorders

For calibrating the optional front and side mounted chart recorders, refer section 6

(Optional Accessory Operation)